Novel Solid-Contact Ion-Selective Microelectrodes for Localized Potentiometric Measurements
authors Lamaka, SV; Taryba, MG; Zheludkevich, ML; Ferreira, MGS
nationality International
journal ELECTROANALYSIS
author keywords Solid contact; pH-selective microelectrode; Mg-selective microelectrode; SIET; Response time; Ionophores; Sensors
keywords SCANNING ELECTROCHEMICAL MICROSCOPY; IUPAC RECOMMENDATIONS; NEUTRAL CARRIER; RESPONSE-TIME; ELECTRODES; MEMBRANE; SENSORS; FLUXES
abstract The design and properties of novel type of solid-contact ionophore-based ion-selective microelectrodes are reported. The microelectrode is based on an insulated needle-shaped metallic wire with an exposed apex. The ion-to-electron transducer is made of poly(3-octylthiophene-2,5-diyl) and placed between an ion-selective membrane and the metallic tip. The ion-selective polyvinyl chloride-based membrane is deposited atop the layer of conductive polymer. The length of the ion-sensitive part of the electrode is less than 10 mu m. pH and Mg2+-selective microelectrodes were constructed and tested showing stable potential and fast response that are essential properties for the practical application of microelectrodes for localized scanning measurements.
publisher WILEY-V C H VERLAG GMBH
issn 1040-0397
year published 2009
volume 21
issue 22
beginning page 2447
ending page 2453
digital object identifier (doi) 10.1002/elan.200900258
web of science category Chemistry, Analytical; Electrochemistry
subject category Chemistry; Electrochemistry
unique article identifier WOS:000272390100007
  ciceco authors
  impact metrics
journal analysis (jcr 2017):
journal impact factor 2.851
5 year journal impact factor 2.469
category normalized journal impact factor percentile 64.407
dimensions (citation analysis):
altmetrics (social interaction):



 


Apoio

1suponsers_list_ciceco.jpg