resumo
The physical, chemical and biological properties of the bioglass reinforced yttria-stabilized composite layer on Ti6Al4V titanium substrates were investigated. The Ti6Al4V substrate was deposited with yttria stabilized zirconia - YSZ as the base layer of thickness approximate to 4-5 mu m, to inhibit metal ion leach out from the substrate and bioglass zirconia reinforced composite as the second layer of thickness approximate to 15 mu m, which would react with surrounding bone tissue to enhance bone formation and implant fixation. The deposition of these two layers on the substrate was carried out using the most viable electrophoretic deposition (EPD) technique. Biocompatible yttria-stabilized zirconia (YSZ) in the form of nano-particles and sol gel derived bioglass in the form of micro-particles were chosen as precursors for coating. The coatings were vacuum sintered at 900 degrees C for 3 h. The biocompatibility and corrosion resistance property were studied in osteoblast cell culture and in simulated body fluid (SBF) respectively. Analysis showed that the zirconia reinforced bioglass bilayer system promoted significant bioactivity, and it exhibited a better corrosion resistance property and elevated mechanical strength under load bearing conditions in comparison with the monolayer YSZ coating on Ti6Al4V implant surface. (C) 2013 Elsevier B.V. All rights reserved.
palavras-chave
OSTEOBLAST-LIKE CELLS; HYDROXYAPATITE COATINGS; TITANIUM; SURFACE; ALLOY
categoria
Materials Science
autores
Ananth, KP; Suganya, S; Mangalaraj, D; Ferreira, JMF; Balamurugan, A
nossos autores
agradecimentos
The authors are grateful to DST-SERB (DST-SERB/F/0849/2012-2013, dated 16.05.2012) for the financial support.